Matrix-assisted laser desorption/ionization (MALDI) is a soft ionization technique used in mass spectrometry, allowing the analysis of biomolecules (biopolymers such as proteins, peptides and sugars) and large organic molecules (such as polymers, dendrimers and other macromolecules), which tend to be fragile and fragment when ionized by more conventional ionization methods. It is most similar in character to electrospray ionization both in relative softness and the ions produced (although it causes many fewer multiply charged ions).

The ionization is triggered by a laser beam (normally a nitrogen laser). A matrix is used to protect the biomolecule from being destroyed by direct laser beam and to facilitate vaporization and ionization.

Contents


Matrix

UV MALDI Matrix List
Compound Other Names Solvent Wavelength (nm) Applications
2,5-dihydroxy benzoic acid[1] DHB, Gentisic acid acetonitrile, water, methanol, acetone, chloroform 337, 355, 266 peptides, nucleotides, oligonucleotides, oligosaccharides
3,5-dimethoxy-4-hydroxycinnamic acid[2][3] sinapic acid; sinapinic acid; SA acetonitrile, water, acetone, chloroform 337, 355, 266 peptides, proteins, lipids
4-hydroxy-3-methoxycinnamic acid[2][3] ferulic acid acetonitrile, water, propanol 337, 355, 266 proteins
?-cyano-4-hydroxycinnamic acid[4] CHCA acetonitrile, water, ethanol, acetone 337, 355 peptides, lipids, nucleotides
Picolinic acid[5] PA Ethanol 266 oligonucleotides
3-hydroxy picolinic acid[6] HPA Ethanol 337, 355 oligonucleotides

The matrix consists of crystallized molecules, of which the three most commonly used are 3,5-dimethoxy-4-hydroxycinnamic acid (sinapinic acid), ?-cyano-4-hydroxycinnamic acid (alpha-cyano or alpha-matrix) and 2,5-dihydroxybenzoic acid (DHB). A solution of one of these molecules is made, often in a mixture of highly purified water and an organic solvent (normally acetonitrile (ACN) or ethanol). Trifluoroacetic acid (TFA) may also be added. A good example of a matrix-solution would be 20 mg/mL sinapinic acid in ACN:water:TFA (50:50:0.1).

Notation for cinnamic acid substitutions.
Notation for cinnamic acid substitutions.
The identity of suitable matrix compounds is determined to some extent by trial and error, but they are based on some specific molecular design considerations:

  • They are of a fairly low molecular weight (to allow facile vaporization), but are large enough (with a low enough vapor pressure) not to evaporate during sample preparation or while standing in the spectrometer.
  • They are acidic, therefore act as a proton source to encourage ionization of the analyte.
  • They have a strong optical absorption in the UV, so that they rapidly and efficiently absorb the laser irradiation.
  • They are functionalized with polar groups, allowing their use in aqueous solutions.

The matrix solution is mixed with the analyte (e.g. protein-sample). The organic solvent allows hydrophobic molecules to dissolve into the solution, while the water allows for water-soluble (hydrophilic) molecules to do the same. This solution is spotted onto a MALDI plate (usually a metal plate designed for this purpose). The solvents vaporize, leaving only the recrystallized matrix, but now with analyte molecules spread throughout the crystals. The matrix and the analyte are said to be co-crystallized in a MALDI spot.

Laser

Matrix-assisted laser desorption ionization depicted with matrix in blue and analyte in red
Matrix-assisted laser desorption ionization depicted with matrix in blue and analyte in red

Lasers Used for MALDI
Laser Wavelength (nm) Reference
Nitrogen laser 337 (Tanaka 1988)[7]
Nd:YAG 355, 266 (Karas 1985)[8]
Er:YAG 2940 (Overberg 1990)[9]
CO2 10,600 (Overberg 1991)[10]

The laser is fired at the crystals in the MALDI spot. The matrix absorbs the laser energy and it is thought that primarily the matrix is ionized by this event. The matrix is then thought to transfer part of its charge to the analyte molecules (e.g. protein), thus ionizing them while still protecting them from the disruptive energy of the laser. Ions observed after this process consist of a neutral molecule [M] and an added or removed ion. Together, they form a quasimolecular ion, for example [M+H]+ in the case of an added proton, [M+Na]+ in the case of an added sodium ion, or [M-H]- in the case of a removed proton. MALDI is capable of creating singly-charged ions, but multiply charged ions ([M+nH]n+) can also be created, as a function of the matrix, the laser intensity and/or the voltage used. Note that these are all even-electron species. Ion signals of radical cations can be observed eg. in case of matrix molecules and other stable molecules.

AP-MALDI

Atmospheric pressure (AP) matrix-assisted laser desorption/ionization (MALDI) is an ionization technique (ion source) that in contrast to vacuum MALDI operates at normal atmospheric environment.[11] The main difference between vacuum MALDI and AP-MALDI is the pressure in which the ions are created. In vacuum MALDI, ions are typically produced at 10 mTorr or less while in AP-MALDI ions are formed in atmospheric pressure. Disadvantage of the AP MALDI source is the limited sensitivity observed and the limited mass range.

AP-MALDI is used in mass spectrometry (MS) in a variety of applications ranging from proteomics to drug discovery fields. Popular topics that are addressed by AP-MALDI mass spectrometry include: proteomics, DNA/RNA/PNA, lipids, oligosaccharides, phosphopeptides, bacteria, small molecules and synthetic polymers, similar applications as available also for vacuum MALDI instruments.

The AP-MALDI ion source is easily coupled to an ion trap mass spectrometer[12] or any other MS system equipped with ESI (electrospray ionization) or nanoESI source.

Mass spectrometer

Sample target for a MALDI mass spectrometer
Sample target for a MALDI mass spectrometer
The type of a mass spectrometer most widely used with MALDI is the TOF (time-of-flight mass spectrometer), mainly due to its large mass range. The TOF measurement procedure is also ideally suited to the MALDI ionization process since the pulsed laser takes individual 'shots' rather than working in continuous operation. MALDI-TOF instruments are typically equipped with an "ion mirror", deflecting ions with an electric field, thereby doubling the ion flight path and increasing the resolution. Today, commercial reflectron TOF instruments reach a resolving power m/?m of well above 20'000 FWHM (full-width half-maximum, ?m defined as the peak width at 50% of peak height).

History

The term matrix-assisted laser desorption ionization (MALDI) was coined in 1985 by Franz Hillenkamp, Michael Karas and their colleagues.[13] These researchers found that the amino acid alanine could be ionized more easily if it was mixed with the amino acid tryptophan and irradiated with a pulsed 266 nm laser. The tryptophan was absorbing the laser energy and helping to ionize the non-absorbing alanine. Peptides up to the 2843 Da peptide melittin could be ionized when mixed with this kind of ?matrix?.[14] The breakthrough for large molecule laser desorption ionization came in 1987 when Koichi Tanaka of Shimadzu Corp. and his co-workers used what they called the ?ultra fine metal plus liquid matrix method? that combined 30 nm cobalt particles in glycerol with a 337 nm nitrogen laser for ionization.[15] Using this laser and matrix combination, Tanaka was able to ionize biomolecules as large as the 34,472 Da protein carboxypeptidase-A. Tanaka received one-quarter of the 2002 Nobel Prize in Chemistry for demonstrating that, with the proper combination of laser wavelength and matrix, a protein can be ionized.[16] Karas and Hillenkamp were subsequently able to ionize the 67 kDa protein albumin using a nicotinic acid matrix and a 266 nm laser.[17] Further improvements were realized through the use of a 355 nm laser and the cinnamic acid derivatives ferulic acid, caffeic acid and sinapinic acid as the matrix.[18] The availability of small and relatively inexpensive nitrogen lasers operating at 337 nm wavelength and the first commercial instruments introduced in the early 1990s brought MALDI to an increasing number of researchers.[19] Today, mostly organic matrices are used for MALDI mass spectrometry.

Use

In Biochemistry

In proteomics, MALDI is used for the identification of proteins isolated through gel electrophoresis: SDS-PAGE and two-dimensional gel electrophoresis. One method used is peptide mass fingerprinting by MALDI-MS, or with post ionisation decay or collision-induced dissociation (further use see mass spectrometry).

In Organic Chemistry

Some synthetic macromolecules, such as catenanes and rotaxanes, dendrimers and hyperbranched polymers, and other assemblies, have molecular weights extending into the thousands or tens of thousands, where most ionization techniques have difficulty producing molecular ions. MALDI is a simple and rapid analytical method that can allow chemists to analyze the results of such syntheses and verify their results.

Reproducibility and performance

The sample preparation for MALDI is important for the result. Inorganic salts which are also part of protein extracts interfere with the ionization process. The salts are removed by solid phase extraction or washing the final target spots with water. Both methods can also remove other substances from the sample. The matrix protein mixture is not homogenous because the polarity difference leads to a separation of the two substances during crystallization. The spot diameter of the target is much larger than that of the laser, which makes it necessary to do several laser shots at different places of the target, to get the statistical average of the substance concentration within the target spot. The matrix composition, the addition of trifluoroacetic acid and formic acid, delay between laser pulses, delay time of the acceleration power, laser wavelength, energy density of the laser and the impact angle of the laser on the target are among others the critical values for the quality and reproducibility of the method.

See also

References

Bibliography

External links

da:MALDI de:Matrix-unterstützte Laser-Desorption/Ionisation es:MALDI-TOF fr:Désorption-ionisation laser assistée par matrice it:MALDI nl:MALDI ja:??????????????????? ru:????????-?????????????? ???????? ?????????/????????? (MALDI) ur:????? ??????? ???? ??????/?????





Source: Wikipedia | The above article is available under the GNU FDL. | Edit this article



Related Links in Matrix-assisted laser desorption/ionization

Search for Matrix-assisted laser desorption/ionization in Tutorials
Search for Matrix-assisted laser desorption/ionization in Encyclopedia
Search for Matrix-assisted laser desorption/ionization in Dictionary
Search for Matrix-assisted laser desorption/ionization in Open Directory
Search for Matrix-assisted laser desorption/ionization in Store
Search for Matrix-assisted laser desorption/ionization in PriceGig



Help build the largest human-edited directory on the web.
Submit a Site - Open Directory Project - Become an Editor

Advertisement

Advertisement



Matrix-assisted laser desorption/ionization
Matrix-assisted_laser_desorption/ionization top Matrix-assisted_laser_desorption/ionization

Home - Add TutorGig to Your Site - Disclaimer

©2008-2009 TutorGig.com. All Rights Reserved. Privacy Statement